DISPOSAL

Discarded items should be cleaned and disinfected prior to

disposal to avoid cross-contamination. It is recommended to

recycle them according to local regulations.

VIGILANCE

In case of any serious incident that has occurred in relation to

the use of the device, please report the incident to CMT at

contact@coremedtech.dk. You can also report it to the competent

authority of the state in which the user and/or patient is

established.

CONTACT INFORMATION

Contact your reseller for requests regarding support,

customer service, software download or an upgrade.

In case of malfunction, please report to your reseller or

CMT at contact@coremedtech.dk.

LEGAL MANUFACTURER

CoreMedTech ApS

Gydevang 35

3450 Allerød

Danmark

Contact: +45 22660442

UK REP

Internatikonal Associates Limited

38 Queen Street, Glasgow,

G1 3DX, UK

UKRP@ia-uk.com

If the scan post has wedged itself and is difficult to remove the

situation can be mitigated by introducing a slight momentum

(rotate).

WARNING

Avoid applying excessive torque to scan post

during retraction as this may cause unnecessary

deformation or even break the Scan Post.

WARNING

It is not recommended to use tools during the

mounting, retraction or dislodging as this may

cause damage, deformation or even break the Scan

Post.

STEP 5: CLEANING AND STORAGE

Following use, the scan posts must be cleaned prior to storage.

Cleaning procedure

• Clean the scan posts by rinsing with clean, preferably purified,

water until dust or debris is removed.

• Dry the articles with a clean lint-free cloth

• Store in appropriate container.

WARNING

The scan posts must be thoroughly cleaned

immediately after use.

STORAGE

The system storage box allows for sorting of scan posts by length

and diameter.

The scan posts kit should be stored away from

direct sunlight.

LIFETIME

The lifetime of the product is dependendant on several

factors and it is recommended that the condition of the scan

post is inspected prior to every use. The maximum amout of

resprocessing cycles is secondary to the how progressed the

general wear is.

Maximum 100 reprocessing cycles

WARNING

The scan posts may be damage and or change

dimensions due to excessive wear and it is

recommended that the Scan Posts are inspected

prior to use and discarded at the first sign of wear

or deterioration.