UK
STEP 2: PREPARATION FOR IN-MODEL USE
Clean the scan body and the associated screwdriver for use.
CAUTION
The scan body and the analog connection must be
undamaged and free from cast material.
CAUTION
Replace the scan body if any damage, i.e. scratches,
deformation, excessive wear, etc., is detected on
either scan body or system screwdriver.
STEP 3: MOUNTING THE SCAN BODY ON THE IMPLANT
ANALOG IN THE GYPSUM MODEL
Place the scan body on the clean implant analog and rotate
until the interface locks and no further rotational movement is
possible. Use the supplied screwdriver to fix the screw finger-
tight.
IMPORTANT NOTICE
To avoid damage and unnecessary wear, the
recommended mounting torque is 5 N·cm.
CAUTION
The screwdriver does not have torque limitation!
STEP 4: SCANNING WITH THE SCAN BODY IN PLACE
Scan the scan body following the scanner’s instructions for use.
Ensure that the entire surface of the scan body is scanned fully
and without surface gaps, especially the head of the scan body,
both top and sides.
STEP 5: UNMOUNTING THE SCAN BODY
Use the supplied screwdriver to unmount the scan body.
STEP 6: CLEANING AND STORAGE
Clean the scan body and screwdriver, if needed, and place them
back in the storage box.
CAUTION
The scan bodies and associated screwdrivers must
be thoroughly cleaned immediately after use.
FADING
IMPORTANT NOTICE
Towards the end of the scan bodies’ lifespan, a slight
change in the surface appearance can be detected.
The fading is purely visual and has no impact on
product functionality or scan quality.
CAUTION
Replace the scan body if any damage, i.e. scratches,
deformation, excessive wear etc. is detected on
either the scan body or the system screwdriver.
STEP 4: SCANNING WITH THE SCAN BODY IN PLACE
Scan the scan body following the scanner’s instructions for use.
Ensure that the entire surface of the scan body is scanned fully
and without surface gaps, especially the head of the scan body,
both top and sides.
WARNING
Position the patient in such a way that, should the
scan body be dropped or come loose nevertheless,
it will not fall into the patient’s respiratory or
digestive tract.
WARNING
To avoid aspiration or swallowing, always have an
assistant in place with suction equipment ready
for use. In case of swallowing of the scan body
or screwdriver, follow standard procedures and
contact the nearest medical facility.
STEP 5: UNMOUNTING THE SCAN BODY
Use the screwdriver to unmount the scan body.
WARNING
Refer to the IMPORTANT NOTICE, CAUTIONS and
WARNINGS in STEP 3 above.
STEP 6: CLEANING AND STORAGE
Clean the scan body and screwdriver, if needed, and place them
back in the storage box.
WARNING
The scan bodies and associated screwdrivers must
be thoroughly cleaned immediately after use.
IN MODEL USE – STEP BY STEP
STEP 1: SELECTION OF THE SCAN BODY
Select the scan body that matches the implant system and
platform.
IMPORTANT NOTICE
Each scan body and screwdriver contain human-
readable markings with a code indicating the
corresponding implant platform and system. The
screwdrivers contain a 2-character system-specific
alphanumeric code, and the scan bodies have an
additional 3rd character referring to the connection
platform.
CAUTION
To avoid damage to the implant analog and
ensure that correct scan data is obtained, it is
recommended that an in-depth knowledge of
preferred systems is obtained, enabling a quick
confirmation of both the connection and the
screwdriver.
CAUTION
Replace the scan body if any damage, i.e. scratches,
deformation, excessive wear, etc., is detected on
either the scan body or the system screwdriver.
CoreMedTech